Welch Allyn Connex Vital Signs Monitor Calibration/Verification Module Notes - Quick Reference Guide User Manual
Page 6

Code Definition /
Description
Origin / Alert
type
Traditional code
map
Display message to
operator
Host controller
action
8
Too few steps.
Patient alert.
C05
Unable to determine
blood pressure. Check
connections; restrict
patient movement.
Make the error
clearable, and
allow another
cycle to be run.
9
A neonate cuff was
detected in adult
mode.
Operator alert. Not applicable.
Message displayed
depends on the host
controller environment.
Message could be due to
neonatal patient
movement.
Make the error
clearable, and
allow another
cycle to be run.
Change patient
mode or an
equipment alert
(C03).
10
Re-inflation too late. Patient alert.
C05
Unable to determine
blood pressure. Check
connections; restrict
patient movement.
Make the error
clearable, and
allow another
cycle to be run.
11
Too many re-
inflates.
Patient alert.
C05
Unable to determine
blood pressure. Check
connections; restrict
patient movement.
Make the error
clearable, and
allow another
cycle to be run.
12
Could not bleed
pressure below
SVRP.
Equipment
alert.
C02
Unable to release cuff
pressure.
Check tubing for kinks
and connection integrity.
Make the error
clearable, and
allow another
cycle to be run.
13
Air leak detected.
Equipment
alert.
C03
Inflation too quick. Check
for kinks in the tubing
and connections.
Make the error
clearable, and
allow another
cycle to be run.
14
Safety check failed. Equipment
alert.
C10
Safety fault detected
during cycle. Cycle
cancelled.
Power down.