beautypg.com

Welch Allyn Connex Vital Signs Monitor Calibration/Verification Module Notes - Quick Reference Guide User Manual

Page 4

background image

Species

Code Definition /

Description

Original / Alert

type

Traditional code

map

Details

Host

controller

action

6

6

Equipment alarm E11

Cuff pressure

duration

exceeded.

Reset Mod PG.

6

7

Equipment alarm Internal error

Internal NIBP

error. Contact

Welch Allyn

Technical

Service.

Power down.

6

8

Equipment alarm Internal error

Internal NIBP

error. Contact

Welch Allyn

Technical

Service.

Power down.

6

9

Host controller

alert

Internal error

None

Exit STAT

mode.

6

11

Host controller

alert

Internal error

None

None

6

12

Host controller

alert

Internal error

None

None

6

13

Host controller

alert

Internal error

None

None

6

22

Equipment alarm Internal error

Internal

malfunction.

Contact Welch

Allyn Technical

Service.

Power down.

This is only

available

during

software

loading.

6

23

Equipment alarm Internal error

Internal

malfunction.

Contact Welch

Allyn Technical

Service.

Power down.

This is only

available

during

software

loading.

6

14

Host controller

alert

Internal error

None

None

6

15

Host controller

alert

Internal error

None

None

6

16

Host controller

alert

Internal error

None

None

6

17

Host controller

alert

Internal error

None

None

6

18

Host controller

alert

Internal error

None

None