beautypg.com

Periodic maintenance, Troubleshooting – Welch Allyn Multi-Operatory Air Abrasion Accessory - User Manual User Manual

Page 16

background image

990229 REV C

16

8. Replace the 4 knurled thumb screws through the pinch tube blocks and tighten

until the pinch tube top blocks are pulled down and meet the bottom pinch tube
blocks under the thumb screws.

CAUTION: When installing new pinch tubes, ensure the pinch tube top blocks
are pulled down completely under the installation screws. Failure to tighten these
screws completely could result in improper blocking of powder, resulting in
air/abrasive release from the inactive handpiece.

9. Reconnect power to the Multi-Operatory Accessory, then power on the Mach

5/5+.

10. Resume normal operation.

Periodic Maintenance

Your Multi-Operatory Accessory requires minimal maintenance. However, Kreativ
recommends that pinch tubes and tubing are checked for wear whenever maintenance
is performed on your Mach 5/5+ system. It is also recommended that both pinch
tubes be replaced as a part of a preventive maintenance program at least once every
three (3) months.

WARNING: Failure to periodically replace the pinch tubes on the recommended
schedule could result in permanent damage to the working components of the Multi-
Operatory Accessory, rendering it unusable.

Troubleshooting

The only user-serviceable parts in the Kreativ Multi-Operatory Accessory are the
pinch tubes. The following table is intended to assist in communicating with your
Welch Allyn / Kreativ service representative.

CONDITION:

THINGS TO DO:

Displays do not illuminate

Check that Multi-operatory Accessory power

cord is plugged in.

Check power supply connector is plugged into the

Multi-Operatory Accessory.

Ensure that all interconnecting cables are

connected per the installation instructions.