Lamp Replacement

1.

Turn unit off and unplug the power cord from both the wall outlet and
the Insight Instruments Light Source.

2.

Remove the lamp door from the bottom of the Insight Instruments
Light Source by unscrewing the lamp access knob on the back of the
lightbox.

3.

Disconnect the lamp connector assembly and remove the lamp from
the holding spring assembly.

4.

Replace with Insight Instruments replacement lamp P/N: 81502 only.
Do not touch the lamp itself, or the interior reflective surface of the
lamp. Skin oils on the lamp will cause premature lamp failure. Hold
lamp by the outside of the reflector or the connector only. Remove
any grease or fingerprints with a clean cotton swab moistened with
alcohol. Do not leave any lint on the lamp.

CAUTION: Lamp is hot. Allow the lamp to cool before
replacing.