Welch Allyn Insight Instruments Solarc Light Source - User Manual User Manual
Insight instruments solarc™ light source
Table of contents
Document Outline
- Insight Instruments Solarc™ Light Source
- For P/N: 20002.000
- Symbol Descriptions
- Indications
- Contraindications
- System Description
- Set-up And Operation
- 1. Plug the Insight Instruments Light Source into a hospital-grade receptacle.
- 2. Turn the light-intensity control knob to its minimum position. Plug the fiber optic bundle int...
- 3. Turn the power switch to the on position. The lamp will take 15 seconds to complete its warm-u...
- 4. Set light intensity to desired level.
- 5. Turn unit off when not in use.
- Mounting
- Lamp Replacement
- 1. Turn unit off and unplug the power cord from both the wall outlet and the Insight Instruments ...
- 2. Remove the lamp door from the bottom of the Insight Instruments Light Source by unscrewing the...
- 3. Disconnect the lamp connector assembly and remove the lamp from the holding spring assembly.
- 4. Replace with Insight Instruments replacement lamp P/N: 81502 only. Do not touch the lamp itsel...
- 5. Reconnect the lamp to the lamp connector and place in the spring mechanism. Verify that the pi...
- 6. Reattach the lamp door securely. The unit will not power up unless the lamp door is seated pro...
- Fuse Replacement
- 1. Turn the unit off and remove the power cord from both the wall outlet and the Insight Instrume...
- 2. Use a small screwdriver to remove the fuse holder above the power cord connector by releasing ...
- 3. Pull out the fuse holder.
- 4. Remove and replace with new fuses of the same type. There is no required orientation of the fu...
- 5. Reinsert the fuse holder by pressing it into the power connector until it snaps into place.
- 6. If the new fuse blows, immediately contact Insight Instruments Technical Service at 1-407-324-...
- Cleaning
- 1. Prior to cleaning, turn power off and disconnect the power cord from both the wall outlet and ...
- 2. Clean the external surfaces by wiping with a cloth dampened with mild soap and water or by usi...
- 3. Wipe clean the power cord by using the same cleaning/disinfecting solutions listed above. Wipe...
- 4. Do not plug the unit into a wall outlet until it is thoroughly dry.
- Specifications
- Troubleshooting
- Maintenance And Repair
- Service
- Insight Instruments, Inc.
- 5400 S. Bryant Ave
- Sanford, FL 32773
- Tel: (800) 255 8354 or (407) 324-2727
- Insight Instruments, Inc.
- 5400 S. Bryant Ave.
- Sanford, FL 32773, USA
- Tel.: 1-800-255-8354 or
- 1-407-324-2727
- Fax: 1-407-324-2744