Specifications, Abpm 6100 hardware manual – Welch Allyn ABPM 6100 Hardware - User Manual User Manual

Page 21

background image

ABPM 6100 Hardware Manual

DIR 80012720 Ver. E

21 / 28


6. Specifications

This section provides the unit specifications for the ABPM 6100 ambulatory blood pressure
monitor.
Power Requirements:

Two “AA” alkaline batteries or high capacity rechargeable batteries (NimH).

Dimensions:

124 x 70 x 33 mm (approximate).

Weight:

270g (including batteries, approximate).

Environmental Operating Condition:

Atmospheric Pressure Range of 700 HPA to 1060 HPA

Protection against electric shock:

Internally powered, Type BF

Operating Ambients:

Temperature: 10°C (50°F) to 50°C (122°F).

Humidity: 20% to 95% RH non-condensing.

Storage Ambients:

Temperature: -20°C (-68°F) to 70°C (158°F).

Humidity: 15% to 95% RH non-condensing.

Altitude: -170 to 1,700 meters

Gauging Ranges:

Systolic Blood Pressure: 60 to 250 mmHg

Diastolic Blood Pressure: 25 to 200 mmHg

Maximum Inflation: 270 mmHg

Heart Rate: 40 to 200 bpm

Memory:

Up to 250 readings using Alkaline batteries

Up to 110 reading using rechargeable NimH

Method of Measurement:

Oscillometric with step deflation.

Accuracy:

± 3 mmHg

International Standards:

EN 60601-1-1:2001 ; Medical electrical equipment

– Part 1-1: General requirements for

safety. Collateral standard: Safety requirements for medical electrical systems
Specific requirements for the protection against hazards of ignition of flammable
anesthetic mixtures and to control the risk of fire (Fire prevention)

EN 60601-2-30:2000 ; Medical electrical equipment

– Part 2-30: Particular requirements

for the safety, including essential performance, of automatic cycling non-invasive blood
pressuring monitoring equipment

EN 60601-1-2:2001 ; Medical Electrical Equipment

– Part 1-2: General requirements for

safety. Collateral Standard: Electromagnetic compatibility

– requirements and tests

EN 1060-1:1996 ; Specification for Non-invasive sphygmomanometers

– Part 1.

General requirements

EN 1060-3:1997 ; Non-invasive sphygmomanometers

– Part 3: Supplementary

requirements for electro-mechanical blood pressure measuring systems

AAMI SP 10 ES1 Category C: 1992 (battery powered) Electronic or automated
sphygmomanometers

Calibration:

Minimally, once per year

Safety System:

Maximum Inflation pressure limited to 300 mmHg.

Auto safety release valve for power failure.

Maximum BP measurement time limited to less than 180 seconds.

Data Connector: