4. Maintenance and Cleaning

Preventative maintenance should be routinely performed to ensure the safe and efficient
operation of the ABPM 6100 monitor. In addition, the monitor should be cleaned after each
use.

Maintenance

The following inspections of the ABPM 6100 unit should be performed on a regular basis:

The PC connection cable should be inspected for any cracks, exposed wires or other
damage.

The monitor itself should be visibly inspected for any signs of damage.

Pneumatic tubing should be inspected for any cracks, fraying, or kinks.

Do not remove any covers or break the warranty seal while inspecting the unit.


If any signs of damage are detected, do not use the ABPM 6100 monitor. It should be sent
to an authorized Welch Allyn service center. See the Service and Warranty section of this
manual for a list of authorized Welch Allyn service centers.

Cleaning

CLEANING THE ABPM 6100 UNIT


The ABPM 6100 unit should be cleaned after every use. Use a soft, damp cloth to remove
any dirt and dust from the unit.

CLEANING AND DISINFECTING THE CUFF

1. Medical Disinfectant

Spray

The cuff may be sprayed with a mild disinfectant solution
(e.g. Cidezyme

®

, ENZOL

®

, or 10% bleach solution. Follow

the direction on the solution), rinsed with distilled water, and
line dry. Ensure that no liquid enters the bladder tubing.

2. Medical Disinfect Wipe

Use a mild disinfectant wipe (e.g. Sani-Cloth

®

) and

thoroughly wet cuff surface and line dry.

3. Machine Wash

Remove bladder to machine wash the cuff shell. Machine
wash warm (50 - 130° F, 10 - 54° C) with a mild detergent
and line dry.

IMPORTANT: The ABPM 6100 cannot be sterilized. DO NOT immerse the
monitor in any fluid or use liquid detergents, cleaning agents, or
solvents to clean it. If the unit is immersed in any liquid, do not use the
unit again. The unit should be sent to an authorized Welch Allyn service
center.