Directions for Use

Chapter 4 Alarms and Alerts

57

Nurse Call

The monitor can be connected to a Nurse Call system through a customized cable that
connects to the Nurse Call connector. When the cable is connected and operational, the
monitor immediately notifies the Nurse Call system when a patient alarm occurs.

To connect the monitor to a Nurse Call system, you must have a cable (Welch Allyn part
number 008-0634-XX or equivalent) that has been adapted to your Nurse Call system. This
cable can be built or customized by a biomedical engineering department or other
appropriate technical resource, using the specification described on

“Nurse Call”

on

page 79.

WARNING Although the Nurse Call option enables remote notification of an
alarm condition, it is not intended to replace appropriate bedside patient
surveillance by trained clinicians.

Note

When a patient alarm occurs, pressing

silences the monitor alarm tone and

the Nurse Call alarm for 90 seconds, but does not suspend the visual alarm
indicator on the monitor.

Nurse Call Cable