Directions for Use

Chapter 3 Patient Monitoring

37

Monitoring SpO

2

Warnings and Cautions — SpO

2

WARNING Always follow the manufacturer’s instructions for care and use of the
SpO

2

sensor.

WARNING The accuracy of the SpO

2

measurement can be affected by any of

the following:

the presence of significant amounts of dysfunctional hemoglobin, such as
carboxyhemoglobin or methemoglobin

the presence of concentrations of some intravascular dyes, sufficient to
change the patient’s usual arterial pigmentation

patient movement

patient conditions such as shivering and smoke inhalation

painted nails

poor oxygen perfusion

anemia or low concentrations of hemoglobin

hypotension or hypertension

severe vasoconstriction

shock or cardiac arrest

venous pulsations or sudden and significant changes in pulse rate

proximity to an MRI environment

moisture in the sensor

excessive ambient light, especially fluorescent

wrong sensor or sensor too tight

WARNING If there is any question of the accuracy of an SpO

2

measurement,

verify the measurement using another clinically accepted measurement method.