Directions for use
Welch Allyn standard laryngoscope blade assemblies
5
Pre-vacuum autoclave sterilization
Packaging: Package the blade assembly in an appropriately sized single pouch or in a
wrap labeled as appropriate for the pre-vacuum autoclave exposure parameters outlined
below; package the blade assembly per facility practice in preparation for autoclave.
Pre-vacuum autoclave sterilization: Follow equipment manufacturer and facility
procedures in the set-up and operation of autoclave equipment. Pre-vacuum autoclave
settings are as follows:
•
Temperature: 132 C (270 F)
• Exposure
time:
4
minutes (wrapped)
• Minimum
dry
time:
20
minutes
Storage: Store per facility practice to allow the packaged blade assembly to remain
clean, dry and ready for service.
Steris
®
Amsco
®
V-PRO
®
low temperature sterilization
Packaging: Package the blade assembly per facility practice in an appropriately sized
single pouch or in a wrap labeled as appropriate for the Steris
®
Amsco
®
V-PRO
®
collection of low temperature sterilization equipment and exposure parameters.
Steris
®
Amsco
®
V-PRO
®
low temperature sterilization: Follow equipment manufacturer
and facility procedures in the set-up and operation of Steris
®
Amsco
®
V-PRO
®
low
temperature sterilization equipment. V-PRO
®
sterilization settings are as follows:
•
V-PRO 1: Lumen cycle
•
V-PRO 1 PLUS: Lumen cycle, non-Lumen cycle
•
V-PRO maX: Lumen cycle, non-Lumen cycle, Flexible cycle
Storage: Store per facility practice to allow the packaged blade assembly to remain
clean, dry and ready for service.
STERRAD
®
hydrogen peroxide gas plasma sterilization
Packaging: Package the blade assembly per facility practice in an appropriately sized
single pouch, tray or in a container labeled as appropriate for the STERRAD
®
hydrogen
peroxide gas plasma exposure parameters.
STERRAD
®
hydrogen peroxide gas plasma sterilization: Follow equipment manufacturer
and facility procedures in the set-up and operation of the STERRAD
®
hydrogen peroxide
gas plasma sterilization equipment. STERRAD
®
hydrogen peroxide gas plasma
sterilization systems and cycles are as follows:
•
STERRAD 100S: Standard (short) cycle
•
STERRAD 100S: Long cycle (outside of US only)
•
STERRAD NX: Standard, Advanced cycles
•
STERRAD 100NX: Express, Standard cycles
Storage: Store per facility practice to allow the packaged blade assembly to remain
clean, dry and ready for service.
End of reprocessing instructions for sterilization.