Directions for use

Welch Allyn standard laryngoscope blade assemblies

5



Pre-vacuum autoclave sterilization

Packaging: Package the blade assembly in an appropriately sized single pouch or in a
wrap labeled as appropriate for the pre-vacuum autoclave exposure parameters outlined
below; package the blade assembly per facility practice in preparation for autoclave.

Pre-vacuum autoclave sterilization: Follow equipment manufacturer and facility
procedures in the set-up and operation of autoclave equipment. Pre-vacuum autoclave
settings are as follows:

Temperature: 132 C (270 F)

• Exposure

time:

4

minutes (wrapped)

• Minimum

dry

time:

20

minutes

Storage: Store per facility practice to allow the packaged blade assembly to remain
clean, dry and ready for service.

Steris

®

Amsco

®

V-PRO

®

low temperature sterilization

Packaging: Package the blade assembly per facility practice in an appropriately sized
single pouch or in a wrap labeled as appropriate for the Steris

®

Amsco

®

V-PRO

®

collection of low temperature sterilization equipment and exposure parameters.

Steris

®

Amsco

®

V-PRO

®

low temperature sterilization: Follow equipment manufacturer

and facility procedures in the set-up and operation of Steris

®

Amsco

®

V-PRO

®

low

temperature sterilization equipment. V-PRO

®

sterilization settings are as follows:

V-PRO 1: Lumen cycle

V-PRO 1 PLUS: Lumen cycle, non-Lumen cycle

V-PRO maX: Lumen cycle, non-Lumen cycle, Flexible cycle

Storage: Store per facility practice to allow the packaged blade assembly to remain
clean, dry and ready for service.

STERRAD

®

hydrogen peroxide gas plasma sterilization

Packaging: Package the blade assembly per facility practice in an appropriately sized
single pouch, tray or in a container labeled as appropriate for the STERRAD

®

hydrogen

peroxide gas plasma exposure parameters.

STERRAD

®

hydrogen peroxide gas plasma sterilization: Follow equipment manufacturer

and facility procedures in the set-up and operation of the STERRAD

®

hydrogen peroxide

gas plasma sterilization equipment. STERRAD

®

hydrogen peroxide gas plasma

sterilization systems and cycles are as follows:

STERRAD 100S: Standard (short) cycle

STERRAD 100S: Long cycle (outside of US only)

STERRAD NX: Standard, Advanced cycles

STERRAD 100NX: Express, Standard cycles

Storage: Store per facility practice to allow the packaged blade assembly to remain
clean, dry and ready for service.

End of reprocessing instructions for sterilization.