Directions for use

Welch Allyn standard laryngoscope blade assemblies

3


Manual cleaning
process:













Drying:


Maintenance,
Inspection and Testing
prior to disinfection or
sterilization:

1. Immerse all components in a neutral pH enzymatic

cleaning solution as directed by solution
manufacturer’s instructions.

2. While components are immersed in the cleaning solution,

use a soft-bristled brush to clean all component surfaces
(the blade, retaining screw, and light carrier) until all
visible contamination is removed.

3. Thoroughly rinse all components in one of the following:

potable filtered water, softened water, or per cleaning
solution manufacturer instructions, or per facility
instructions to remove cleaning solution.

After cleaning, dry all components with a clean cloth or allow
to air dry.

1. Inspect each component area (per Figure 3) for damage or

deterioration. Also, inspect the following areas for
deterioration:

• hooks

of

laryngoscope

blade

(1) and light assembly

(2)

• lamp

(3)

• threads

of

blade

(1) and locking screw (4)

WARNING: Discard any component that shows

evidence of damage or deterioration.

Contact Welch Allyn for component replacement.

2. Reassemble the blade assembly by reversing steps

illustrated in Figure 2.

3. Attach blade assembly to a clean and disinfected test

handle in known working condition. Verify that:

Blade assembly engages and locks onto handle.

Blade assembly deploys into its locked position on
handle AND lamp illuminates.

• Verify

that

light

output

is

satisfactory.

If the lamp fails to light or output is low, check or
replace the lamp.

After cleaning; choose either cold solution high level disinfection method (page 4)
for any blade or one of the four sterilization methods (page 4-5) except REF 68470

Figure 3

1. Laryngoscope blade

2. Light assembly
3. Lamp
4. Locking screw