beautypg.com

Reprocessing instructions – Welch Allyn Standard laryngoscope blade assemblies - User Manual User Manual

Page 2

background image

2

Directions for use

Welch Allyn s t a n d a rd laryngoscope blade assemblies


Reprocessing instructions

Reprocessing refers to procedures for cleaning and disinfection or sterilization of devices.
The standard laryngoscope blade assemblies must be reprocessed prior to first use and
between each use using any of the following methods outlined in this document:

Cleaning and cold solution high level disinfection

Cleaning and gravity autoclave sterilization

Cleaning and pre-vacuum autoclave sterilization

Cleaning and Steris

®

Amsco

®

V-PRO

®

low temperature sterilization

Cleaning and STERRAD

®

hydrogen peroxide gas plasma sterilization


Welch Allyn has validated these instructions as being capable of preparing its laryngoscope
blade for re-use. The user must ensure that the reprocessing as actually performed by the
user's personnel, with the user's equipment and materials, achieves the desired result. This
may require validation and routine monitoring of the user's actual process.

Cleaning instructions

CAUTION: Do not remove lamp during reprocessing. Damage to device could occur.


Point of use:

1. Separate blade assembly from handle and place the blade

assembly into suitable containment for subsequent
reprocessing. See Figure 1. Do not place the blade
assembly with sharp devices.

2. Prevent the blade assembly from drying per facility

practice (for example, use immersion, neutral pH
enzymatic pre-cleaner or foam, or cover with a moist
towel).

Preparation for
decontamination:

1. Disassemble the blade assembly per Figure 2.

2. Keep all blade components together to prevent loss.

3. Select a neutral pH enzymatic cleaner.

4. Prepare cleaning solution per cleaning solution

manufacturer instructions.













Figure 1


Figure 2