Results

Reliability Number indicates the number of good readings obtained and their
consistency, based on a 1 to 9 scale. The higher the number, the better.

>6 is typically acceptable

5 is marginal / repeat the test if possible

<4 is poor / repetition of the test is necessary

If the patient did not appear to fixate on the unit during data acquisition, repeat

the test, even if reliability numbers are >6.

S is the sphere in diopters.

C is the cylinder power in diopters.

A +9.99 or -9.99 indicates a reading outside the unit’s measurement range.

The SW X.XXe after the date on the printout is the software revision level in
the SureSight unit.

Sphere

Cylinder

Reliablity
Number

Axis

Child/Adult
Mode