beautypg.com

Welch Allyn SureSight Autorefractor - User Manual User Manual

Page 21

background image

17

If you are at the correct distance and not acquiring data:

ensure you are level and square with the patient,

tell the patient to look into the red light,

slowly move the cross-hairs around the eye, and

make sure the patient’s eyelids are not occluding the pupil

aim higher when over-refracting through glasses

To stop a test at any time, hold any button until 5 tones sound.

When the unit is off, any button wakes the unit and displays the previous
results.

To Print: Press the print button until beep sounds. Keep unit aimed at printer
during beep.

Do not send 2

nd

printout to printer before 1

st

printout has completed printing