4 printouts, 5 output format – Rice Lake Ranger Count 3000 Series, Ohaus User Manual
Page 35

Ranger™ Count 3000 Series EN-33
7.4 Printouts
The following sample print outs are generated by the Print button, “P” Command or alternate
print command. The content of the printout is defined in the Print Content menu item. A
maximum of 24 characters can be printed on each line.
Description
Comment
PN: 123456789926
Part Number line
If Printx Content P/N is ON
363 PCS NET
Result line
If Printx Content Result is ON
0.1420 kg G
Gross value line
If Printx Content Gross is ON and a tare value is entered
0.0420 kg NET
Net value line
If Printx Content Net is ON and a tare value is entered
0.1000 kg PT
Tare value line
If Printx Content Tare is ON and a tare value is entered
APW: 0.11500 g
APW line
If Printx Content APW is ON
N: 3
Accumulation line
If Printx Content Accu is All
ACC: 956 Pcs
Accumulation line
If Printx Content Accu is All
7.5 Output Format
Weight string print format:
Field
Weight space Unit
space Stability space G/N
Space Term.
Char(s)
Length 9
1
5
1
1
1
NET
1
Each field is followed by a single delimiting space (ASCII: 32)
Definitions:
Weight
- Up to 9 characters, right justified, - at immediate left of most significant character
(if negative).
Unit
- Up to 5 characters, left justified. If the Unit in the Print Content menu was set to
OFF, the unit will be removed in the weight string and replaced by spaces.
Stability
- “?” character is printed if not stable. If weight is stable neither “?” or following
space is printed.
G/N
- “NET” printed if weight is net weight, ‘G’ or nothing printed if weight is a gross
weight.
Terminating Character(s) - terminating character(s) printed depending on FEED menu setting.