Front panel calibration – Rice Lake RailBoss Rail Scales - Calibration Guide User Manual
Page 4

4
Railboss Calibration Guide
Front Panel Calibration
The CALIBR submenu (under the SCALES menu, see
Figure 6) is used to calibrate the
920i
. The zero, span,
and linear calibration point displays provide a set of
softkeys used specifically for calibration procedures:
+/–
Toggles to allow entry of negative or
positive values
Last Zero
Recalls the last established zero value to
allow calibration without removing tests
weights or product from scale.
Calibrate
Performs calibration for the selected
point
Temp Zero
Temporarily zeroes the displayed weight
o f a n o n - e m p t y s c a l e . A f t e r s p a n
calibration, the difference between the
temp zero and the previously calibrated
zero value is used as an offset.
Millivolts
(or
Counts
)
Toggles between display of captured
A/D counts and captured millivolts
values; allows entry of calibration values
in mV or counts
Figure 6. WZERO Calibration Display
To calibrate the indicator using the front panel, do the
following:
1. Ensure the J-Box trimming procedure has been
successfully completed.
2. Place the indicator in setup mode by removing
t h e l a rg e f i l l i s t e r- h e a d s c r e w f r o m t h e
enclosure. Insert a screwdriver or similar tool
into the access hole and press the setup switch
once. The indicator display changes to show
scale configuration menus.
3. With the
SCALES
menu highlighted, press the
down
key, then select the Scale 1. Press
down
again (
GRADS
parameter highlighted), then
press
left
to highlight the
CALIBR
submenu (see
Figure 5). Press
down
to go to zero calibration
(
WZERO
). See Figure 6.
4. Ensure scale reading is at zero, then press
down
again to show the current WZERO value, then
press the
Calibrate
softkey to calibrate zero.
When complete, the new A/D count for the
zero calibration is displayed. Press
Enter
to
save the zero calibration value and go to the
next prompt (
WVAL
).
5. With
WVAL
displayed, press
down
to show the
stored calibration weight value. Use the
numeric keypad to enter the actual value of the
RailBoss Calibrator load to be used (30,000 lb.
is recommended), then press
Enter
to save the
value and go to span calibration (
WSPAN
).
6. Use the RailBoss Calibrator to place a 30,000
lb. load on the scale. Press
down
again to show
the current WSPAN value, then press the
Calibrate
softkey to calibrate span. When
complete, the new A/D count for the span
calibration is displayed. Press
Enter
again to
save the span calibration value and go to the
next prompt (
WLIN
).
7. When complete, press the
Save and Exit
softkey
to exit setup mode.
03/01/2011
01:57PM
Calibrate
WZERO
WVAL
REZERO
1291
Deadload Calibration
+/-
Millivolts
Zero Count =>
Last Zero
Temp Zero
CALIBR
SCALE1
SCALES