Legal-for-trade sealing, Save, Weight calibration – Rice Lake Intercomp CS 1500 - Users Manual User Manual
Page 22: Finish

CS1500, Users
Rev S, June, 2009
22
If your capacity is:
Set your graduation to:
300 lb / 150 kg
.1 lb / 0.05 kg (9)
500 lb / 250 kg
0.2 lb / 0.1 kg (8)
1000 lb / 500 kg
0.5 lb / 0.2 kg (7)
2000 lb / 1000 kg
1 lb / 0.5 kg (6)
5000 lb / 2500 kg
2 lb / 1 kg (5)
10000 lb / 5000 kg
5 lb / 2 kg (4)
20000 lb / 10000 kg
10 lb / 5 kg (3)
Save
•
At this point the scale saves any changes that have been made. This allows
changes to be made to “EE-10
EE-10
EE-10
EE-10” through “EE-14
EE-14
EE-14
EE-14” without having to do a complete
calibration. The scale can be turned off and any changes so far will be saved.
Weight calibration
20. The scale shows “LL-00
LL-00
LL-00
LL-00”. With no weight on hook simultaneously press the
ZERO and lb/kg keys. This reads the pad zero. On the next screen, enter
the scale's capacity.
21. The scale shows “LL-01
LL-01
LL-01
LL-01”. Apply the first weight. With the first weight stable
on the hook, simultaneously press the ZERO and lb/kg keys. Then enter the
value of the applied weight.
22. The scale shows “LL-02
LL-02
LL-02
LL-02”. Apply the second weight. With the second weight
stable on the hook, simultaneously press the ZERO and lb/kg keys. Then
enter the value of the applied weight.
23. The scale shows “LL-03
LL-03
LL-03
LL-03”. Apply the third weight. With the third weight stable
on the hook, simultaneously press the ZERO and lb/kg keys. Then enter the
value of the applied weight.
Finish
The new calibration information is saved.
24. Return the calibration blocking switch to it's original position (RUN). This
prevents accidental entry into the calibration mode.
25. Verify the calibration.
26. Calibration complete.
Legal-for-Trade Sealing
1. On the front of the CS1500, replace the 2 upper right screws with the drilled screws
provided.
2. Thread a lead & wire seal through these 2 screws.
3. Crimp the lead seal tightly.