8

The HI-4422/FP2000/FP4000/FP5000 probe measures field
strength in each of three axes. It performs a vector addition
calculation on the readings and sends the result to the receiver
via a fiber optic cable. Data from each axis can be viewed
individually, or can be combined. Frequency response of the
HI-4422/FP2000/FP4000/FP5000 is 10 kHz to 1 GHz; dynamic
range is 1 to 300 Volts/meter (V/m).

SPECIFICATIONS

Dynamic Range

1 to 300 Volts/meter (V/m)

Ranges

10, 30, 100, 300 Volts/meter full scale

Frequency
Response

10 kHz to 250 MHz ± 0.5 dB

250 MHz to 1.0 GHz ± 1.0 dB

Linearity

± 0.5 dB full scale (F.S.): ± 2 least
significant bits (LSBs) of A/D converter

Isotropicity

± 0.5 dB

Overload Withstand

1000 Volts/meter maximum, all ranges

Environmental
Operating Temp

10

°C to 40 °C

(+50

°F to +104 °F)

Environmental
Humidity

5% to 95% relative humidity, non-
condensing

Fiber Optic Cable
Connector

Standard FSMA

Battery

3.6 VDC, 1400 mA-h rechargeable
Nickel-Cadmium (NiCd)

Battery Charger

115/230 VAC, approximately 1 hour

Probes

64 mm (2.5 in) cube with probe shields
on three sides

Probe Mount

¼ - 20 UNC tapped hole (internal
thread) in base of probe

Weight

0.4 Kg (14 oz.)

Optional Equipment

See Table 2