Archived 4/2/10 – ETS-Lindgren FP5000 Isotropic Electric Field Probe (Archived) User Manual
Page 8
Isotropic Electric Field Probe
HI-4422/FP2000/FP4000/FP5000
© ETS-LINDGREN L.P. – July, 2005
Revision G- Part # H-600054
8
The HI-4422/FP2000/FP4000/FP5000 probe measures field
strength in each of three axes. It performs a vector addition
calculation on the readings and sends the result to the receiver
via a fiber optic cable. Data from each axis can be viewed
individually, or can be combined. Frequency response of the
HI-4422/FP2000/FP4000/FP5000 is 10 kHz to 1 GHz; dynamic
range is 1 to 300 Volts/meter (V/m).
SPECIFICATIONS
Dynamic Range
1 to 300 Volts/meter (V/m)
Ranges
10, 30, 100, 300 Volts/meter full scale
Frequency
Response
10 kHz to 250 MHz ± 0.5 dB
250 MHz to 1.0 GHz ± 1.0 dB
Linearity
± 0.5 dB full scale (F.S.): ± 2 least
significant bits (LSBs) of A/D converter
Isotropicity
± 0.5 dB
Overload Withstand
1000 Volts/meter maximum, all ranges
Environmental
Operating Temp
10
°C to 40 °C
(+50
°F to +104 °F)
Environmental
Humidity
5% to 95% relative humidity, non-
condensing
Fiber Optic Cable
Connector
Standard FSMA
Battery
3.6 VDC, 1400 mA-h rechargeable
Nickel-Cadmium (NiCd)
Battery Charger
115/230 VAC, approximately 1 hour
Probes
64 mm (2.5 in) cube with probe shields
on three sides
Probe Mount
¼ - 20 UNC tapped hole (internal
thread) in base of probe
Weight
0.4 Kg (14 oz.)
Optional Equipment
See Table 2
Archived 4/2/10