Isotropic Electric Field Probe

HI-4422/FP2000/FP4000/FP5000

© ETS-LINDGREN L.P. – July, 2005

Revision G- Part # H-600054

10

XMIT/RCV

The fiber optic cable assembly from the receiver is attached to
the probe via two connectors. The cable ends are color-coded—
white for XMIT, yellow for RCV. Identically colored dots are
located on the probe housing adjacent to these connectors. Be
sure that each cable is attached to the proper probe connector.

Figure 2: Switch and Connectors

When the cables are not attached, always cover the probe
connectors with the protective plastic covers supplied with the
unit, or with similar material. This prevents dirt and other
contaminants from entering the connector, causing
communication problems.

ARM/OFF

The ARM/OFF switch activates and deactivates the probe. In
the ARM position, its internal 3.6 VDC NiCd battery powers the
probe: in the OFF position, the probe is inactive. To prolong
battery life, set this switch to OFF when the probe is not in use.

CHARGER

A standard fast charger is supplied with the
HI-4422/FP2000/FP4000/FP5000. When charging is complete,
the fast charger acts as a trickle charger. Battery performance
will not degrade, regardless of how long it is on the trickle
charger.