4.0
MAINTENANCE
This section explains which maintenance tasks can be
performed by the user. It also provides information regarding
replacement and optional parts. If you have any questions
concerning probe maintenance, consult ETS-Lindgren Customer
Service.
MAINTENANCE RECOMMENDATIONS
Maintenance of the HI-4422/FP2000/FP4000/FP5000 is limited
to external components such as cables or connectors.
Any calibration or maintenance task that requires probe
disassembly should be performed at the factory. Check with
ETS-Lindgren Customer Service (+1-512-531-6400) before
opening the unit to avoid problems with your probe's warranty.
NOTE: Opening the probe enclosure may void your
warranty. If your system is under warranty, contact ETS-
Lindgren Customer Service before performing any
maintenance inside the probe.
PROBE SHIELDS
Over time, the probe shields (the white, cone-shaped structures
with the axis labels) accumulate foreign materials; in particular,
bits of dirt or metal may become embedded in the shield surface.
Shields must be kept clean to maintain proper probe operation.
NOTE: NEVER attempt to clean the shields while they
are mounted on the probe housing.