Loader rear panel controls, A- fuse, B- voltage select – CDI Torque SURETEST Torque Calibration System User Manual
Page 63: C- ac power input, D- solenoid control

MANUAL LOADER
CHAPTER 6
6-62
Loader Rear Panel Controls
The 2000-600-02 Manual Loader rear panel is shown in Figure 6-3.
Figure 6-3: 2000-600-02 Manual Loader Rear Panel Controls
A- Fuse
The loader power supply is protected with a 1 amp slo-blow fuse.
B- Voltage Select
A selector switch allows the loader to use both 110VAC or 220VAC.
C- AC Power Input
The loader uses either 110VAC or 220VAC power. The loader connects to the SURETEST Base
Unit by the 2000-117 cable.
D- Solenoid Control
AC power is applied to the loader to operate a pair of safety solenoids used to protect the
transducers from overloading. Should the applied torque exceed 110% of the transducer capacity,
in either CW or CCW directions, the Base Unit energizes the solenoids to lock the hand crank.