Calibrate tester – CDI Torque SURETEST Torque Calibration System User Manual
Page 55

CALIBRATION
CHAPTER 5
5-54
Use DOWN Cursor key to scroll down to Calibration Item. Press ‘ENTER’.
The display reads:
4.
Enter the following code into the display using the Cursor Keys:
5111
( Left / Right to select digit ; Up / Down = Increment / Decrement value ; ‘Enter’ to accept)
Calibrate tester
Using a laboratory millivolt meter, measure the voltage across pin 2 (ground) and 19 (Gage
Excitation) of the calibration fixture.
The
display
reads:
Example:
3.0211 volts
Program the measured voltage into the SURETEST Base Unit using the Cursor Keys. (Use Up,
Down, Left and Right Cursor Keys to match the Suretest display to the millivolt-meter).
Then Press ‘Enter’ to accept.
5.
Next, the display reads:
Short pins 5 and 6 together on the calibration fixture. Then hit Enter key.
6.
Next, the display reads:
Remove the short between pins 5 and 6 of the calibration fixture.