Calibration cw – CDI Torque SURETEST Torque Calibration System User Manual
Page 51

CALIBRATION
CHAPTER 5
5-50
6.
Enter the following code into the display using the Cursor Keys:
5222
( Left / Right to select digit ; Up / Down = Increment / Decrement value ; ‘Enter’ to accept)
Calibration CW:
7. Select
Calibrate
CW. Press ‘ENTER’.
A warning is displayed.
Select ‘
Yes
’. Press ‘
Enter
’ .
8. Using the Up or Down Cursor keys, enter “
1
” for a
single transducer
.
Using the Up or Down Cursor keys, enter “
4
” for the
4-in-1 transducer.
9.
Press ENTER. The display reads: