Printer serial output port, Serial communications, Analog output – CDI Torque SURETEST Torque Calibration System User Manual
Page 23: Loader control relays, Display capacity, Units of measurement, Accuracy, Range and resolution, Temperature

FUNCTIONAL DESCRIPTIONS AND SPECIFICATIONS
CHAPTER 2
Printer Serial Output Port
Default: 9600 Baud Rate.
Serial Communications
RS-232 (true)
300—19.2K Baud
8 data bits
1 stop bit
no parity
Computer Serial COM Port (OPTIONAL)
Default: 19,200 Baud Rate.
RS-232 (True)
Analog Output
+(CW)/-(CCW) 1.8V at transducer full range linearity
±1% of reading
Loader Control Relays
Two, normally open, Form A, rated 12VDC @ 1/2A close contact at 110% CW or CCW of torque
transducer range. For additional information refer to Transducer and System specifications.
Display Capacity
(16 bit A/D), 5 digits, ±32,000 counts
Units of Measurement
Torque
ft-lb, in-lb, in-oz, Nm, dNm, cNm, mkg, cmkg
Torque Resolution
Refer to transducer range and display resolution tables in Functional Description of this chapter.
Accuracy
±0.2% of reading @ 25°C, within specified range, when used as prescribed with the 2000-600-02
Loader.
Range and Resolution
Refer to transducer range and display resolution tables in Functional Description of this chapter.
Temperature
Operating Temperature
10 to 32°C (50 to 90°F)
Storage
-20 to 50°C (-2 to 122°F)
Humidity
Up to 90% non-condensing
Temperature Drift
+0.02%/°C (+.011%/°F)