Calibration and parameter setting – Salter Brecknell SBI-505 User Manual
Page 7
![background image](/manuals/634853/7/background.png)
If the weight is stable, press” PRINT” you can print the weight.
COUNT
1.In weighing mode, load the Pieces on the platform scales, Press”
Count” the indicator show” PCS 0” press” Zero” key in the quantity,
press” Print” to confirm it
2. Load the goods on the platform scales, the indicator will show the
quantity.
1. Press” Count” to return to weighing mode.
2. If you want to weigh different goods, at weighing mode, put the
sample on the platform scales, press” Count” the indicator show”0”
Press “Count” hold it and then press “ON/OFF” the indicator show”
PCS 0”, press “Zero” input the sample quantity, press “Print” to
confirm it. Then repeat the step 2 and 3.
Calibration and Parameter setting
Enter setting
There have two methods to enter the setting menu:
1. when the switch “ CAL” is off, press the “PRINT” hold it and
then press” Total” this enters the C08-39 settings.
2. Take out the sealing screw on the back of indicator, move the
sealing switch to second position. Press “PRINT” hold it and then
press “Total” key this enters C01-C39 settings.
Rea
Inside View, Calibration Switch