3= close display if no change within 3min.

5= close display if no change within 5 min.
LCD Version:
0=Close he backlight
1= backlight when the weight change or

press the keyboard

2=constant backlight

Hold

function

N.A

SBI-505

C11
Hold mode

option

:0=close hold function

1=Peak hold /2=Data Hold

Instruction:
Peak-hold: it shows the max. data,

mainly application for materials testing,
such as tension and pulling force.

Date-hold: it shows current weight value.

Mainly application for animal weighing.

Kg/lb

conversion

C12
Kg/lb
conversion


C12=0 stop kg/lb conversion
C12=1 kg/lb conversion is ok

Upper/lower

limit alarm



C13
Upper limit
alarm value


You can set it within the max. capacity limit


C14

Lower limit

alarm value

Inner Code

display

C15
Check inner
code


enter C15 to check the inner code

Date and time

C16
Date

Enter C16, you can set the date,
from left to right: year/month/day

C17
Time

Enter C17, you can set the time from
left to right: hour/min./sec.