xviii

PARAMETER

(Soft Key Text)

DESCRIPTION

AVAILABLE

CHOICES

F1

Graduations

Specifies number of full-scale graduations. Value should be

consistent with legal requirements and environmental limits on

the useful system resolution.

1,000 2,000

5,000 10,000

24,000 30,000

50,000

F3

Zero Track

Selects the range within which the scale will automatically zero.

Note that the scale must be in standstill to automatically zero.

Selections are in Display Divisions.

0d

0.5d

√√√

1d

3d

5d

F5

Motion Band

Sets the level at which motion is detected by comparing the

present display update with the previous one. If motion is not

detected for two seconds or more, scale is in standstill and can

process a Print or Zero command. Maximum value varies

depending on local regulations.

1d

√√√

3d

5d

10d

F9

Display

Divisions

Determines the desired weight increments. Value should be

consistent with legal requirements.

1

2

5

F10

Decimal Pt.

Determines location of the decimal point.

0 0.0

0.00 0.000

0.0000

F16

Zero Calibration

Performs zero calibration.

See chapter 6 for calibration

instructions

F17

Span Calibration

Performs scale calibration.

See chapter 6 for calibration

instructions

F22

Creep Tracking

(CrpTrk)

Sets the range within which the scale will attempt to offset the

effects of load cell creep. Use a lower setting for lower capacity

load cells and vice-versa. Selections are in seconds.

FACTORY SET – DO NOT CHANGE

Off

10, 15, 20, 25, 30,

50,

F23

Creep Time

Sets the range within which the scale will attempt to offset the

effects of load cell creep. Use a lower setting for lower capacity

load cells and vice-versa. Selections are in seconds.

FACTORY SET – DO NOT CHANGE

1, 2, 3, 4, 5, 10,

20, 30,

F24

Calibration Adjust

Allows experienced service technicians to adjust calibration to

compensate for linearity adjustments. USE WITH CAUTION