Salter Brecknell 308BP User Manual
Page 18
![background image](/manuals/634832/18/background.png)
xviii
PARAMETER
(Soft Key Text)
DESCRIPTION
AVAILABLE
CHOICES
F1
Graduations
Specifies number of full-scale graduations. Value should be
consistent with legal requirements and environmental limits on
the useful system resolution.
1,000 2,000
5,000 10,000
24,000 30,000
50,000
F3
Zero Track
Selects the range within which the scale will automatically zero.
Note that the scale must be in standstill to automatically zero.
Selections are in Display Divisions.
0d
0.5d√
√√√
1d
3d
5d
F5
Motion Band
Sets the level at which motion is detected by comparing the
present display update with the previous one. If motion is not
detected for two seconds or more, scale is in standstill and can
process a Print or Zero command. Maximum value varies
depending on local regulations.
1d√
√√√
3d
5d
10d
F9
Display
Divisions
Determines the desired weight increments. Value should be
consistent with legal requirements.
1
2
5
F10
Decimal Pt.
Determines location of the decimal point.
0 0.0
0.00 0.000
0.0000
F16
Zero Calibration
Performs zero calibration.
See chapter 6 for calibration
instructions
F17
Span Calibration
Performs scale calibration.
See chapter 6 for calibration
instructions
F22
Creep Tracking
(CrpTrk)
Sets the range within which the scale will attempt to offset the
effects of load cell creep. Use a lower setting for lower capacity
load cells and vice-versa. Selections are in seconds.
FACTORY SET – DO NOT CHANGE
Off
10, 15, 20, 25, 30,
50,
F23
Creep Time
Sets the range within which the scale will attempt to offset the
effects of load cell creep. Use a lower setting for lower capacity
load cells and vice-versa. Selections are in seconds.
FACTORY SET – DO NOT CHANGE
1, 2, 3, 4, 5, 10,
20, 30,
F24
Calibration Adjust
Allows experienced service technicians to adjust calibration to
compensate for linearity adjustments. USE WITH CAUTION