Chapter 7--maintenance, Maintenance, Cleaning the platecrane – Hudson Robotics PlateCrane EX User Manual
Page 38: Change the fuse(s)

PlateCrane EX User Manual
Hudson Robotics, Inc.
34
CHAPTER 7--Maintenance
1. Maintenance
WARNING- -
Qualified personnel can perform all maintenance procedures in this manual. Only
Hudson representatives should perform any maintenance not discussed in this
manual.
Gloves should be worn during any cleaning procedure.
2. Cleaning the PlateCrane
Note: DO NOT use/spray abrasive cleaners onto PlateCrane.
Prior to cleaning, disconnect system power.
Clean external surfaces only; use of a pre-packaged iopropanol wipe is the only
recommended cleaning technique.
Allow unit to dry for at least 10 minutes prior to connecting system power.
Clean finger pads of gripper with alcohol or other residue-free solvent.
Clean any spills on PlateCrane immediately.
Never allow water or fluid to leak inside the PlateCrane.
3. Change the Fuse(s)
Extra fuses are shipped with the PlateCrane. Fuses burn out occasionally and must be
replaced. The following fuses are used with the PlateCrane:
Metric: 1 x 2 amp, 5 x 20 mm Time Delay Fuse
WARNING- -
Be sure the power to the PlateCrane is OFF and the power cord is unplugged from
the instrument or power source before proceeding with the following instructions.
Turn the PlateCrane OFF and remove the power cord from the AC inlet on the back
of Crane.
Using a small flat head screwdriver, gently pry the fuse holder tray away from its
housing to access the fuses (see Figure 17). The Fuse holder door swings down.