Cooper Instruments & Systems MasterLink 3000 Wireless Load Cell Amplifier User Manual
Page 14

CF 187
12
RevA
Items you can change:
Power On Auto Zero
Here you can determine whether the device performs
automatic zero when it is powered on.
Enter zero to disable this function.
If you enter a non-zero value then when the handheld
is first turned on it checks the value read from the
acquisition module. If this falls within ± of this value
then the display will be altered so this reads zero.
In its simplest form you could select two for a linear
calibration. For more complex calibrations which
include linearization select three to nine points.
Zero Indication Band
Using this setting you can mask tiny changes in input
after you press the Tare button.
Entering zero will disable this function.
Entering a non-zero value will provide a band within
which the display will always read zero.
Once the reading exceeds this value the real weight
will be displayed as no taring is taking place.