Bio-Rad Aurum™ Total RNA 96 Kit User Manual
Page 21

17
Problem
Possible Cause
Recommended Solution
Clogging of RNA
Excessive amount of
Reduce volume of
binding plate
starting material per well culture used
Incomplete mixing of lysis Mix RNA lysis solution
and alcohol solutions
and alcohol thoroughly
by pipetting up and
down
Incomplete
Pipette lysate up and
homogenization of cell
down further to reduce
lysate
lysate viscosity
Incomplete digestion with Increase duration of
lysozyme or lyticase
lysozyme or lyticase
digestion; use fresh
enzyme
Low RNA yield
Low amount of starting
Increase starting material
material
amount up to the
maximum indicated for
the specific starting
material type
Incorrect use of wash
Add the appropriate
solutions
volume of 95–100%
ethanol to the low
stringency wash solution
before initial use
Incorrect preparation
Use only the DNase
of DNase dilution
dilution solution provided
in the kit to dilute the
DNase I
Low sample eluate
See problem “Low or
volume
highly variable eluate
volumes among wells”
Inefficient elution
Preheat the elution
solution to 70°C in water
bath prior to the elution
step
Incomplete lysis
Pipet lysate up and down
further to facilitate lysis