Bio-Rad Aurum™ Total RNA 96 Kit User Manual
Page 19

15
Section 9
Troubleshooting Guide
Problem
Possible Cause
Recommended Solution
Difficulty achieving Purge step in protocol
If sealing of wells with
–17 to –23 inHg
sealing tape restores
negative pressure
negative pressure, no
corrective action is
required; however, do
not leave sealed
Open vacuum regulator
Turn control knob fully
clockwise to close
regulator
Plate not seating on
Press down gently on
vacuum manifold
RNA binding plate to
seat
Manifold top not seated
Ensure that manifold
properly on base
top is seated properly on
base
Residue on gasket
Rinse manifold and
gasket with distilled
water and wipe dry
Gasket detached from
Press down to
manifold
reestablish uniform
adhesive contact or
replace gasket
Worn manifold gasket
Replace gasket
Cross-contamination Abrupt application of
Increase negative
between adjacent
negative pressure during pressure gradually
wells
elution
over 5–10 sec using
the vacuum regulator
Residual wash buffer on
Blot RNA binding plate
drip directors
drip directors with paper
towels; ensure that the
plate underside is dry
- S3e™ Cell Sorter (36 pages)
- S3e™ Cell Sorter (132 pages)
- xMark™ Microplate Absorbance Spectrophotometer (20 pages)
- iMark Microplate Absorbance Reader (112 pages)
- Microplate Manager Software (138 pages)
- iMark Microplate Absorbance Reader (48 pages)
- Model 550, Ultramark, and Benchmark Microplate Reader Accessories (39 pages)
- Model 550, Ultramark, and Benchmark Microplate Reader Accessories (15 pages)
- Model 680 Microplate Reader Accessories (150 pages)
- Model 550, Ultramark, and Benchmark Microplate Reader Accessories (31 pages)
- Model 680 Microplate Reader Accessories (68 pages)
- Mouse Typer Isotyping Kit (14 pages)
- ZOE™ Fluorescent Cell Imager (36 pages)
- TC20™ Automated Cell Counter (34 pages)
- ChromLab™ Software (178 pages)
- Design Your NGC™ System (306 pages)
- Design Your NGC™ System (300 pages)
- BioLogic DuoFlow 10 System V9-8 (4 pages)
- BioLogic DuoFlow 10 System HR Gradient Pump Maintenance Kit (4 pages)
- BioLogic DuoFlow 10 System DuoFlow Chromatography System Version 4.0 Software New Features (7 pages)
- BioLogic DuoFlow 10 System DuoFlow Chromatography System Starter Kit (40 pages)
- BioLogic DuoFlow 10 System DuoFlow Chromatography System 5.0 Software New Features (6 pages)
- BioLogic DuoFlow 10 System pH Monitor (4 pages)
- BioLogic DuoFlow 10 System Chromatography System (239 pages)
- BioLogic Mixers (2 pages)
- BioLogic QuadTec™ Detector and Components (58 pages)
- BioLogic Sample Loading Loops (20 pages)
- BioLogic LP System (28 pages)
- BioLogic LP System (5 pages)
- BioLogic LP System (1 page)
- LP Data View Software for BioLogic LP System (5 pages)
- BioLogic LP System (106 pages)
- LP Data View Software for BioLogic LP System (4 pages)
- Model EP-1 Econo Pump (1 page)
- Model EP-1 Econo Pump (23 pages)
- Econo Gradient Pump Kits (19 pages)
- Econo Gradient Pump Kits (51 pages)
- Model EP-1 Econo Pump (27 pages)
- Profinia™ Protein Purification Instrument (224 pages)
- Profinia™ Protein Purification Instrument (4 pages)
- Profinia™ Protein Purification Instrument (2 pages)
- Profinia Protein Purification System Accessories (1 page)
- Profinia Protein Purification System Kits (24 pages)
- Profinia Protein Purification System Kits (24 pages)
- Profinia Protein Purification System Buffers and Reagents (4 pages)