Bio-Rad Profinity IMAC Resins User Manual
Page 6

The cartridges can be used with any liquid
chromatography system capable of setting a high
pressure limit of 45 psi (equivalent to 3 bar or 300 kPa).
Alternatively, luer fittings offer convenient connection
directly to a Luer-Lok syringe for quick, one-step
purification.
Table 1 Bio-Scale Mini IMAC Cartridge
Specifications
Sizes
1 ml and 5 ml bed volumes
Dimensions
1 ml: 40 mm length x 5.6 mm inner
diameter
5 ml: 40 mm length x 12.6 mm
inner diameter
Maximum pressure tolerance
45 psi
Recommended flow rates
1 ml: 1–2 ml/min (240 cm/hr–480 cm/hr)
5 ml:5–10 ml/min (240 cm/hr–480 cm/hr)
Maximum flow rate
1 ml: 6 ml/min (1440 cm/hr)
5 ml: 20 ml/min (963 cm/hr)
Fittings
Female luer-lock inlet and
male luer-lock outlet
Column material
Polypropylene
Frit material
Polyethylene (HDPE)
Shipping conditions
20% ethanol
Storage recommendations
20% ethanol
Autoclavability
Not autoclavable