Bio-Rad Profinity IMAC Resins User Manual
Page 28

Section 11
Troubleshooting Guide
Problem
Possible Cause
Solution
Cartridge clogging
Particulates in samples
Filter all samples and buffers
or slow flow rate
or buffers
through 0.2 µM filter prior to
application
Sample too viscous
Add nuclease to lysate to
degrade DNA. Centrifuge
and filter lysate again
No target protein
Low level of target
Check expression level by
in eluant
protein in starting
SDS-PAGE
material
Target protein not
Check levels of target protein
binding, or eluting in
in lysate, flow-through, wash,
wash fractions
and eluted fractions. Check
for presence of His-tag with
anti-His antibody
Target protein in
His-tag not accessible
Reclone His-tag onto opposite
flow through
terminus (N or C-terminus)
His-tag not accessible
Purify protein under denatur-
ing conditions to expose His-
tag
Proteolysis & removal
Include protease inhibitors in
of tag
lysis buffer (or reaction), or
purify in the cold
25