Bio-Rad Profinity IMAC Resins User Manual
Page 21
Table 6. Purification Method Suggestions
1 ml Cartridge
5 ml Cartridge
Step
CV
Flow Rate
Flow Rate
Equilibrate
5
2 ml/min
10 ml/min
Lysate Load
5 to 10
2 ml/min*
10 ml/min*
Wash 1
6
2 ml/min
10 ml/min
Wash 2
6
2 ml/min
10 ml/min
Elute
5
2 ml/min
10 ml/min
Standard methods that are compatible with any
type of chromatography system are listed below. To
maximize binding capacity with large proteins
(>100 kD), for purification at 4°C, or for purifications
under denaturing conditions, the lysate load flow
rate* can be decreased (to 0.5 ml/min for the 1 ml
cartridge and 2 ml/min for the 5 ml). This will have
to be determined empirically for individual proteins.
1.
Equilibrate the cartridge with 5 column volumes
(CV) of equilibration/wash buffer 1 at 2 ml/min.
2.
Load the sample lysate at 2 ml/min.
3.
Wash the cartridge with 6 CV of wash buffer
1 at 2 ml/min.
18
10005455A.qxp 3/27/2006 2:42 PM Page 18