Table 3. Troubleshooting Guide
Problem
Cause
Characteristics
High back
1. Bed collapse from excessive Sudden catastrophic pressure
pressure
flow rate
increase at flow above
recommended maximum
2. Chemical contamination
Gradual increase in pressure
during use
3. Trapped air
Column returned to use
after storage
4. Microbial contamination
Column returned to use
after storage
5. Change in ionic form
Rapid pressure increase after
buffer change
6. Bed degradation from
Sudden catastrophic pressure
wrong organic solvent
increase with solvent change
Loss of resolution
1. Particulate contamination
Gradual loss in efficiency or
changing retention
2. Change in ionic form
Changing retention times;
Sudden change which may be
accompanied by poor efficiency
3. Bed collapse from excessive Retention times constant; poor
flow rate
efficiency; skewed peaks.
Sudden change.
4. Bed compression
Retention times constant;
lowered efficiency; tailing
skewed peaks. Gradual change.
5. Bed degradation from wrong Changing retention times
`
organic solvent
accompanied by poor efficiency
6. Excessive dead volume
Inability to match test
chromatogram