Table 1. Specifications and Operating Guidelines for Aminex Columns
Aminex HPX-87C
Aminex HPX-42C
Aminex HPX-87P
Column, 300 x 7.8 mm
Column, 300 x 7.8 mm
Column, 300 x 7.8 mm
Catalog number
125-0095
125-0096
125-0098
Resin ionic form
Calcium
Calcium
Lead
Support
Sulfonated divinyl
Sulfonated divinyl
Sulfonated divinyl
benzene-styrene benzene-styrene benzene-styrene
copolymer
copolymer
copolymer
Particle size
9 µm
25 µm
9 µm
Maximum pressure
1,500 psi
800 psi
1,500 psi
Maximum flow rate
1.0 ml/min
0.7 ml/min
1.0 ml/min
at temperature max.
Maximum temperature 85°C
85°C
85°C
Typical mobile phase
H
2
O
H
2
O
H
2
O
pH range
5–9
5–9
5–9
Guard cartridge*
125-0128
125-0128
125-0119
Fermentation Aminex
Aminex
Monitoring Column,
HPX-72-O column,
HPX-72S column,
150 x 7.8 mm
300 x 7.8 mm
300 x 7.8 mm
Catalog number
125-0115
125-0141
125-0146
Resin ionic form
Hydrogen
Hydroxide
Sulfate
Support
Sulfonated divinyl
Sulfonated divinyl
Quaternized divinyl
benzene-styrene benzene-styrene benzene-styrene
copolymer
copolymer
copolymer
Particle size
9 µm
11 µm
11 µm
Maximum pressure
1,500 psi
1,200 psi
1,200 psi
Maximum flow rate
1.0 ml/min
1.0 ml/min
1.0 ml/min
at temperature max.
Maximum temperature 65°C
35°C
65°C
Typical mobile phase
H
2
O
H
2
O
H
2
O
pH range
1–3
9–14
4–9
Guard cartridge*
125-0129
125-0133
125-0132