Lamp replacement Starter replacement

8.3.2 Starter replacement (P/N 100-1370)

Turn off the power.

Disconnect the power cord from the Universal Hood II.

Remove the four screws located on the left/right sides of the drawer.

Remove the cover with the UV filter by sliding it forward.

Place it down on its backside on a non-abrasive surface.

Caution: Do not put the UV filter directly on the bench.

To remove the starter, rotate it counter-clockwise and pull out.

To replace the starter insert it into the holder and rotate clockwise.

Reassemble the cover and retighten the screws on both sides.