3 connecting the cabling harnesses – Bio-Rad Gel Doc™ XR+ System User Manual
Page 15
Page 15 of 42
Center the lens in the opening of the hood as viewed from
inside of the hood
e) Fix the Camera bracket to the bracket on the Universal Hood II using the screw, washer
and knurled knob.
f) Lower the Camera so that the lens fit properly on the gasket of the black adapter ring
then tighten the black screw on the tripod mount
Upon completion of the Camera installation, refer to Section 4.2.3 and Appendix B for a
complete description of the cable connections.
4.2.3 Connecting the cabling harnesses
The cable connections are different if you are installing a Gel Doc
XR or a ChemiDoc
XRS Camera. Please refer to the instructions included in Appendix A, B respectively for Gel
Doc
XR/ ChemiDoc XRS cable wiring.
The controls for the Motorized Zoom Lens can be operated from the membrane touch pad of
the Universal Hood II or from the PC using the buttons that appear in the window when the
option Gel Doc
XR or ChemiDoc XRS is selected in the File Menu of the Quantity One
program. To use this option it is necessary to connect the Universal Hood II serial port
(DB25) (on the back left side) to the COM1 of the computer. The PC/Mac cable is included.
Note: In the case when a Mac has a USB port instead of the serial port, use the USB to
serial adapter (included with Mac systems) to control the lens via software.