7

Section 3
Technical Information

3.1. Sensitivity with Laser Scanning

When scanned on the Molecular Imager FX system, an SDS-PAGE gel stained
with Flamingo fluorescent gel stain should exhibit a limit of sensitivity in the range
of 0.25–0.5 ng (Fig. 2).

0.5 ng

Fig. 2. Serial 2-fold dilutions of SDS-PAGE molecular weight standards, broad range, on a

Criterion Tris-HCl 4–20% gel. The gel was stained with Flamingo fluorescent gel stain and imaged

with the Molecular Imager FX system. The arrow indicates the lane containing 0.5 ng of each standard.

10003321A.qxp 8/19/2005 11:50 AM Page 7