Electroporation conditions are known, Exponential waveform protocols, Cell number <5 x 10 – Bio-Rad Gene Pulser MXcell™ Electroporation System User Manual
Page 9: Gene pulser mxcell electroporation guide
9
Electroporation Conditions Are Known
Cell number <5 x 10
6
Exponential Waveform Protocols
Use preset protocols
Opt mini 96-well/Exp to identify optimal voltage
and capacitance.
Gene Pulser MXcell
Electroporation Guide
Check all well
sets after each
electroporation
for transfection
efficiency and
cell viability.
Choose the preset
mini protocols that
best match your
known conditions and
edit parameters as
needed, following the
instructions.
1
2
3
4
5
6
A
200 V
0 µF
20 V
0 µF
00 V
0 µF
20 V
200 µF
20 V
0 µF
20 V
00 µF
B
C
D
Would you like to further optimize? For manual programming, go to
Keep the preset values for capacitance
and resistance:
n
Enter the known voltage value
(median voltage) in well set ABCD 2
n
Enter the median voltage decreased
by 0 V in well set ABCD 1
n
Enter the median voltage increased
by 0 V in well set ABCD
n
Enter the median voltage value in well
sets ABCD 4–6
n
Save the protocols under a new name