Additional preset protocols available, Preset protocols application parameters, Plate type # of wells used – Bio-Rad Gene Pulser MXcell™ Electroporation System User Manual
Page 18: Gene pulser mxcell electroporation guide
96-well/Exp
96
96
Use for initial optimal protocol
Applies the same waveform
identification for many cell types
across the whole plate
96-well/Sqr
96
96
Use for initial optimal protocol
Applies the same waveform
identification for many cell types
across the whole plate
Mixed 96-well/Exp, Sqr
96
96
Use for mixing different waveforms
Exp: 20 V, 0 μF
by alternating rows of exponential
Sqr: 20 V, 20 ms
(20 V/0 μF) and square waveforms
(20 V/20 ms)
Mixed 24-well/Exp, Sqr
24
24
Use for mixing different waveforms by
Exp: 20 V, 0 μF, 1,000 Ω
alternating rows of exponential
Sqr: 20 V, 20 ms, 1,000 μF, 1,000 Ω
(20 V/0 μF) and square waveforms
(20 V/20 ms)
24-well/Exp
24
24
Use for initial protocol setup for
Same conditions for the
many cell types
whole plate
24-well/Sqr
24
24
Use for initial protocol setup for
Same conditions for the
many cell types
whole plate
Opt 24-well/Exp, Sqr
24
24
Use with cell line with no protocol
Exp: 10–40 V, 200–1,000 μF
reference; this protocol includes a
Sqr: 10–40 V, –0 ms
range of common starting conditions
Uniform 24-well/Exp, Sqr
24
24
Use with a set of defined conditions
Exp: 20 V, 0 μF, 1,000 Ω
to compare different cell lines and
Sqr: 20 V, 20 ms, 1,000 μF, 1,000 Ω
electroporation of different molecules
within the same or different cell lines
12-well/Exp
12
12
Use for initial protocol setup for many
Same conditions for the
cell types
whole plate
12- well/Sqr
12
12
Use for initial protocol setup for many
Same conditions for the
cell types
whole plate
Opt 12-well/Exp, Sqr
12
12
Use with cell line with no protocol reference; Exp: 10–400 V, 200–00 μF
this protocol includes a range of common
Sqr: 10–00 V, 1–2 ms
starting conditions
Uniform 12-well/Exp, Sqr
12
12
Use with a set of defined conditions to
Exp: 20 V, 0 μF, 1,000 Ω
compare different cell lines and
Sqr: 20 V, 20 ms, 1,000 μF, 1,000 Ω
electroporation of different molecules
within the same or different cell lines
Mixed 12-well/Exp, Sqr
12
12
Use for mixing different waveforms by
Exp: 20 V, 0 μF, 1,000 Ω
alternating rows of exponential
Sqr: 20 V, 20 ms, 1,000 μF, 1,000 Ω
(20 V/0 μF) and square waveforms
(20 V/20 ms)
Additional Preset Protocols Available
Gene Pulser MXcell
Electroporation Guide
Preset Protocols
Application
Parameters
See the Gene Pulser MXcell
Electroporation System
manual for more information.
Plate
Type
# of
Wells
Used
1