beautypg.com

Electroporation conditions are known, Square waveform protocols, Cell number >5 x 10 – Bio-Rad Gene Pulser MXcell™ Electroporation System User Manual

Page 12: Gene pulser mxcell electroporation guide

background image

Electroporation Conditions Are Known

Cell number >5 x 10

6

Gene Pulser MXcell

Electroporation Guide

Check all well

sets after each

electroporation
for transfection

efficiency and

cell viability.

For an overview of

common analytical

methods, go to

pages 15–17.

Visit us on the Web at discover.bio-rad.com

| Known Conditions: Square Setup — Whole Plate

Choose whole plate

preset protocols that

best match your

known conditions

and edit parameters

as needed, following

the instructions.

Use the top half of the electroporation plate to identify optimal voltage.

n

Enter the known capacitance value in well sets ABCD 1–12, or keep preset values
if unknown

n

Enter the known voltage (median voltage) value in well sets ABCD 4–6

n

Vary the voltage by 100 V increments around the median value in other well sets

n

Keep the preset values for all other parameters

Use the bottom half of the electroporation plate to identify optimal pulse duration.

n

Enter the known voltage value in well sets EFGH 1–12

n

Keep the preset values for all other parameters

Save the protocols under a new name.

Square Waveform Protocols

Use whole plate preset protocols

96-well/Sqr, Vgrad, Dgrad to identify

optimal voltage and pulse duration.

Good results? Congratulations! Be rewarded for your work and share your protocol
with scientists worldwide. Visit www.bio-rad.com/genetransferprotocols/.

Would you like to further optimize? For manual programming, go to page 14.

1

2

3

4

5

6

7

8

9

10

11

12

A

DV

(V)

100

100

100

200

200

200

00

00

00

400

400

400

B

C

D

E

DD

(ms)

10

10

10

1

1

1

20

20

20

0

0

0

F

G

H

12