Electroporation conditions are known, Square waveform protocols – Bio-Rad Gene Pulser MXcell™ Electroporation System User Manual
Page 10
Electroporation Conditions Are Known
Cell number <5 x 10
6
Gene Pulser MXcell
Electroporation Guide
Check all well
sets after each
electroporation
for transfection
efficiency and
cell viability.
Square Waveform Protocols
Use preset protocols
Opt mini 96-well/Sqr to identify optimal voltage
and pulse duration.
1
2
3
4
5
6
A
200 V
20 ms
20 V
20 ms
00 V
20 ms
20 V
1 ms
20 V
20 ms
20 V
2 ms
B
C
D
Would you like to further optimize? For manual programming, go to
Keep the preset values for pulse duration,
capacitance, and resistance:
n
Enter the known voltage value
(median voltage) in well set ABCD 2
n
Enter the median voltage decreased
by 0 V in well set ABCD 1
n
Enter the median voltage increased
by 0 V in well set ABCD
n
Enter the median voltage value in well
sets ABCD 4–6
n
Save the protocols under a new name
Square Waveform Protocols
Use preset protocols
Opt 96-well/Sqr, NP, D to identify optimal number of pulses.
1
2
3
4
A
20 ms
1 P
1 ms
2 P
10 ms
2 P
7 ms
P
B
C
D
Keep the preset values:
n
Enter the voltage that yielded the
best results from the
Opt mini
96-well/Sqr experiment
n
Save the protocols under a new name
For further optimization of your experiment, identify the optimal number of pulses.
Choose the preset
mini protocols that
best match your
known conditions and
edit parameters as
needed, following the
instructions.
10