LaMotte 5 Series TDS Meter User Manual
Page 28
Single-Point Calibration
A single-point calibration refers to calibrating at one
conductivity value that will be used for all 5 conductivity
ranges.
By selecting [no] to single-point calibration, a calibration
for each conductivity range can be performed.
1. Enter Advanced Set-up.
2. Press
or
button until [S.P.CA] appears on the
LCD. Press ENTER button.
3. Press
or
button to select either [Yes] or no].
4. Press ENTER button to select. The meter will return
to the menu, [S.P.CA].
5. Press
or
button to move to the next menu or
press CAL to exit to the measurement mode.
Restore Factory Default Values
This function allows all parameters to be reset to the factory default settings.
This clears all calibration data and any other set-up
functions that may have been changed.
IMPORTANT: Once activated the meter settings and
calibration data will be erased. Always exercise caution.
Meter reset is not reversible.
1. Enter Advanced Set-up.
2. Press
or
button until [UrSt] appears on the LCD.
Press ENTER button.
3. Press
or
button to select either [Yes] or [no].
4. Press ENTER button to select.
5. The meter will return to the measurement mode after
the turn-on initialization.