Temperature Coefficient
The temperature coefficient is the amount of change in
conductivity per degree of temperature; it is expressed in
percent per °C. Entering the exact temperature
coefficient of the solution being measured will accurately
compensate the temperature for almost any solution. The
temperature coefficient can be set between 0.0 and 3.0 %
per °C. The meter default is 2.1% per °C.
6. Enter Advanced Set-up.
7. Press

or

button until [t.Co %] appears on the

LCD. Press ENTER button.

8. Press

or

button to select a value between 0.0 and

3.0.

9. Press ENTER button to select. The meter will return

to the menu, [t.Co %].

Press

or

button to move to the next menu or press CAL to exit to the

measurement mode.

Normalization Temperature
The meter can be set to normalize the conductivity measurements to a
standard temperature of either 25 °C or 20 °C.
The default value is 25 °C.
1. Enter Advanced Set-up.
2. Press

or

button until [t.nr °C] appears on the

LCD. Press ENTER button.

3. Press

or

button to select either [25.0 °C] or [20.0

°C].

4. Press ENTER button to select. The meter will return

to the menu, [t.nr °C].

5. Press

or

button to move to the next menu or

press CAL to exit to the measurement mode.

Press

ENTER

to

c

onfirm

s

election;

o

r

Press

CAL

to

e

scape

without

confirming

s

election

%

%

%

HOLD

ENTER

Press

ENTER

to

c

onfirm

selection;

o

r

Press

CAL

to

e

scape

without

confirming

selection

C

°

C

°

C

°

HOLD

ENTER

27

This manual is related to the following products: