Page 25

background image

Overview of TDS 6 Set-Up Menu

Select Cell Constant
It is possible to select a cell constant of K = 1.0, 10, or 0.1.
Use a cell of K = 1.0 for midrange measurements
Use a cell of K = 10 for high range measurements
(above 20 mS or 10 ppt).
Use a cell of K = 0.1 for low range measurements
(below 20 µS or 10 ppm).
The cell included with the meter has a cell constant
of K = 1.0.
1. Enter Advanced Set up.
2. Press

or

button until [CELC] appears on the

LCD. Press ENTER button.

3. Press

or

button to select either “1.0”, “0.1” or

“10.0”. Ensure the cell constant selected correspond
with the conductivity electrode that is being used
with the meter.

4. Press ENTER button to select. The meter will

return to the menu, [CELC].

5. Press

or

button to move to the next menu or

press CAL to exit to measurement mode.

%