LaMotte Tracer Salt Meter User Manual
Page 21
TROUBLESHOOTING
Problem
Cause
Action
Reading is frozen
HOLD mode
Press MODE/HOLD
button to exit HOLD
mode.
“BAT” message
Batteries low
Replace batteries
Meter will not calibrate
in pH
Clogged or
contaminated reference
junction
Clean junction
Damaged or worn
sensing membrane
Replace electrode
Contaminated pH
buffers
Use fresh buffers
Meter will not calibrate
in conductivity mode
Trapped air bubbles
Tap probe or stir a
sample to release air
bubbles
Dirty probe
Clean conductivity
probe
Damaged probe
Replace probe
Contaminated
conductivity standards
Use fresh Standards
Meter will not turn on
Batteries low or dead
Replace batteries
Barrery polarity
Replace batteries with
correct polarity
“RENEW” message
pH sensor needs
recalibration
Recalibrate meter
with fresh buffers
pH sensor slope has
fallen below
acceptable limits
Recalibrate meter
with fresh buffers
21