LaMotte Tracer Salt Meter User Manual
Page 20
REPLACING THE ELECTRODE
1. Unscrew and remove the electrode collar. Turn collar
counter-clockwise.
2. Gently rock the electrode side to side, while pulling it away from the
meter, until it disconnects from the electrode socket.
3. To attach an electrode, align the slots and carefully plug the
electrode into the meter socket.
4. Firmly tighten the electrode collar to create a seal with the rubber
gasket between the electrode and the meter.
Electrode Cleaning Recommendations
Do not soak the electrode in the solutions for longer than the
recommended length of time. To do so may cause a reference potential
shift which will cause a degradation in performance or failure. When
cleaning the electrode, take care not to scratch or damage the sensing
surface or the platinized electrode surfaces.
Contaminant
Cleaning Solution
Procedure
Water soluble
Deionized water
Soak or scrub with a soft brush.
Recondition in pH 4 or 7 buffer
for 1 hour.
Grease and
oil
Warm water and
household detergant
Soak or scrub with a soft brush,
maximun 10 minutes. Rinse
thoroughly with DI water.
Recondition in pH 4 or 7 buffer
for 1 hour.
Heavy grease
Alcohol
Soak for a maximum of
5minutes. Scrub with a soft
brush. Rinse thoroughly with
DI water. Recondition in pH 4
or 7 buffer for 1 hour.
Lime and
hydroxide
coatings
10% Acetic acid
Soak until coating dissolves,
maximum 5 minutes. Rinse
thoroughly with DI water.
Recondition in pH 4 or 7 buffer
for 1 hour.
20