LaMotte Tracer Salt Meter User Manual

Page 16

background image

4. The meter will automatically recognize and calibrate to the

conductivity standard. The display will briefly indicate “SA” and
“End” and then return to the measurement mode.
NOTE: “SA” will not appear if the calibration fails.

5. The calibration range indicator will appear on the display for each

range that is calibrated during a power on cycle.

NOTE: Each time the calibration mode is entered all calibration
range indicators will be cleared, but only the calibration data for the
currently selected range will be replaced. The calibrations for the
other two ranges will be saved even though the indicators for those
ranges are no longer displayed. Calibration of all three ranges must
be performed during one power on period for all three calibration
range indicators to be displayed.

NOTE: As with all combination pH/conductivity probes, at low
conductivity levels the flow of the pH electrolyte may affect the
readings and result in artificially high conductivity readings. To
eliminate the interference, low level conductivity or TDS
measurements should be taken with a meter that measures
conductivity only.

16

Low Range, 84 µS/cm

Medium Range, 1413 µS/cm

High Range, 12.88 mS/cm (12,880 µS/cm)