beautypg.com

LaMotte 2020 ClearTrace User Manual

Page 29

background image

To view or reassign the
parameter driving the output
(turbidity or temperature) use
the up/down arrows to select
Sensor or Temperature.
Accept the displayed
parameter with the ENTER
key. Exit without changing
by using the ESC key.

To set the 4 milliamp value
select “Set 4 mA Value” with
the up/down arrows and the
ENTER key.

Use the up/down arrows for
fine adjustment of the low
sensor or temperature value.
Use the left/right arrows for
course adjustment. Accept
the displayed value with the
ENTER key. Exit without
changing by using the ESC
key.

To set the 20 milliamp value
select “Set 20 mA Value”
with the up/down arrows and
the ENTER key.

Use the up/down arrows for
fine adjustment of the low
sensor or temperature value.
Use the left/right arrows for
course adjustment. Accept
the displayed value with the
ENTER key. Exit without
changing by using the ESC
key.

28

Output Param?
(Sensor

)

MENU

ESC

ENTER

(Temperature

)

t

t

t

t

Analog Output
Calibrte

MENU

ESC

ENTER

t

t

t

t

t

Analog Output
Set 4 mA Value

MENU

ESC

ENTER

t

t

t

t

t

Set 4 mA Value?
(

0.0) NTU

MENU

ESC

ENTER

t

t

t

t

Analog Output
Set 20 mA Value

MENU

ESC

ENTER

t

t

t

t

t